亚洲欧美日韩国产综合在线,JAPAN少妇洗澡VIDEOS,久久久久久精品天堂无码中文字幕,摄政王被男人C的合不拢腿H男男


免費注冊快速求購


分享
舉報 評價

高低溫兩用恒溫水槽榮計達HWY-30

參考價面議
具體成交價以合同協議為準

該廠商其他產品

我也要出現在這里

本水浴適用于各種試件的養護,控溫范圍在常溫條件以下使用的儀器,具有制冷速度快、噪音低及恒溫精度高等優點。用途廣泛,可與針入度連接做符合各種溫度試驗。是檢測和試驗單位的設備。

詳細信息 在線詢價

一、產品簡介:
本水浴適用于各種試件的養護,控溫范圍在常溫條件以下使用的儀器,具有制冷速度快、噪音低及恒溫精度高等優點。用途廣泛,可與針入度連接做符合各種溫度試驗。是檢測和試驗單位的設備。
1. Product introduction This water bath is suitable for the maintenance of various specimens. The temperature control range of the instrument used under normal temperature conditions has the advantages of fast cooling speed, low noise and high temperature accuracy. It has a wide range of uses and can be connected with the penetration degree to meet various temperature tests. It is the equipment of testing and testing unit.

二、技術參數:
1、設備控溫范圍:3-60℃
2、電源電壓220v士10v
3、顯示精度:±0.1℃
4、水浴尺寸:460X320X300
5、加熱功率:2000w
6、制冷功率:280W
7、總重:63kg
8、外形尺寸長660mm寬650mm高1050mm
2. Technical parameters 1. Equipment temperature control range: 3-60℃ 2. Power supply voltage 220v ± 10v 3. Display accuracy: ±0.1℃ 4. Water bath size: 460X320X300 5. Heating power: 2000w 6. Cooling power: 280W 7. Total weight: 63kg 8. Dimensions: length 660mm, width 650mm, height 1050mm

三、安裝說明:
1、設備要安放置在室溫10℃到30℃;相對濕度不應高于80%的室內。室內應有良好的通風。
2、拆箱時要注意不要磕碰箱體及部件。
3、本設備電源建議用帶漏電保護的空氣開關,
4、本設備應放量在平整的地面,并且要求設備底部的四個足要全部接觸地面并且著力均勻。
3. Installation instructions: 1. The equipment should be placed in a room with a room temperature of 10℃ to 30℃; the relative humidity should not be higher than 80%. The room should have good ventilation. 2. Be careful not to bump the box and parts when unpacking. 3. It is recommended to use an air switch with leakage protection for the power supply of this equipment. 4. The equipment should be placed on a flat ground, and all four feet at the bottom of the equipment should be in contact with the ground and the force should be even.

四、調試說明:
1、接通電源后將水箱內加上水,對設備進行調試,加水水位不能低于水箱內的篦子。(注意:不加水不能開機!)
2、加滿水后請檢查設備各個部位是否有漏水現象。如果有漏水現象請緊固該部位
螺栓或通知本公司
3、開機,依次按下電源鍵=》制冷鍵=》水泵鍵。這時水泵攪拌裝置啟動。但是,壓
縮機不會啟動(說明:壓縮機在出廠時已經設量了延時保護裝置,一般需要5-8分鐘才
啟動)
4、試機溫度為:依次為25℃、15℃、5℃。首先從25℃開始,當溫控表顯示溫度到達
25℃后,加熱系統會自動運行,而制冷系統會始終運轉,靠加熱系統來補償由于制冷系
統帶來的熱量損失,(我公司技術部之所以這樣設計線路原因:其一,為了更精準的控制
溫度,其二,為了保護壓縮機并延長壓縮機的使用年限)。這時水浴內水的溫度還會繼較
下降到大概低于設置溫度3到5℃,然后逐漸上升直至到達所設置的控溫溫度并恒溫。
15℃與5℃的試機過程依次類推。
試機完畢后將水浴內的水放出,并把各個部位擦拭干凈以備后用。
Four, debugging instructions: 1. After the power is turned on, add water to the water tank to debug the equipment. The water level cannot be lower than the grate in the water tank. (Note: You can't boot without adding water!) 2. After filling up with water, please check whether there is water leakage in all parts of the equipment. If there is water leakage, please tighten the part Bolt or notify the company 3. Turn on the machine and press the power button = "cooling button =" water pump button in turn. At this time, the pump stirring device is activated. However, pressure The shrink machine will not start (Note: The compressor has a delay protection device set at the factory, and it usually takes 5-8 minutes to start up) 4. The test machine temperature is: 25°C, 15°C, and 5°C in sequence. First start at 25℃, when the temperature control meter s that the temperature reaches After 25℃, the heating system will operate automatically, while the refrigeration system will always operate. The heating system is used to compensate the refrigeration system. The heat loss caused by the system, (the reason why the technical department of our company designed the circuit in this way: First, for more precise control Temperature, secondly, in order to protect the compressor and extend the service life of the compressor). At this time, the temperature of the water in the water bath will continue to be higher Decrease to about 3 to 5°C below the set temperature, and then gradually increase until it reaches the set temperature control temperature and keeps constant. The test machine process at 15°C and 5°C is analogized in turn. After the test machine is finished, let out the water in the water bath and wipe all parts clean for later use.
volume_up
content_copy
share

五、操作步驟:
1、將預制好的試樣整齊的排列于工作室內部的托板上,向工作室內加滿水,使水
浸沒試件并高出試件30mm左右。然后接通電源依次打開制冷攪和水泵開關。
2、將所需溫度設定在溫控儀表上。設定方法如下:按下SET鍵隨即松開會看到下
排設定顯示部位的數碼管出現閃動狀態。按△增加,按V設定值減小,設定亮畢后,
按SET鍵返回。溫控表將進入自動控制程序。
3、當溫度達到到控溫儀的所需控制溫度時,溫控儀會自動恒溫。此時紅燈亮并
使水箱溫度穩定在該溫度上,此時記下時間。
4、試件在水箱內溫度恒定至少30分鐘后方可進行試驗。
5、試驗結束后,旋動水箱底部的放水閥將箱內水放凈,擦凈儀器表面水份將儀器
放置干燥通風處。
5. Operation steps: 1. Arrange the prefabricated samples neatly on the pallet inside the studio, and fill the studio with water to make the water The specimen is immersed and is about 30mm higher than the specimen. Then turn on the power supply and turn on the cooling stir and water pump switch in turn. 2. Set the required temperature on the temperature control instrument. The setting method is as follows: press the SET button and then loosen the meeting to see the next The digital tube on the display position of the row setting flashes. Press △ to increase, press V to decrease the setting value, after the setting is bright, Press SET to return. The temperature control meter will enter the automatic control program. 3. When the temperature reaches the required control temperature of the temperature controller, the temperature controller will automatically maintain a constant temperature. The red light is on and Make the temperature of the water tank stabilize at this temperature, and write down the time at this time. 4. The test piece can be tested only after the temperature in the water tank is constant for at least 30 minutes. 5. After the test is over, turn the drain valve at the bottom of the water tank to drain the water, and wipe the surface of the instrument to remove the water Place in a dry and ventilated place.
volume_up
content_copy
share

六、注意事項:
1、本產品使用時應該有良好的接地。
2、水浴后蓋距離墻面不能少于50cm。
3、水槽內應該加人含礦物質很少的飲用水,以免水箱及循環管件內結垢。
4、使用電壓應該嚴格控在220V±10V。
5、使用中不可以開關縮機,使用亮畢后依次關閉水泵、制冷、電源開關。
6、使用完畢后將水浴內的水放出,并把各個部位擦拭千凈以各后用。
7、壓縮機在*不使用的情況下很容易出現故障,所以本設備應該每個月進行一
次試運行,以防壓縮機損壞。
6. Matters needing attention: 1. This product should be well grounded when using it. 2. The distance between the back cover of the water bath and the wall shall not be less than 50cm. 3. Drinking water with few minerals should be added to the water tank to avoid scaling in the water tank and circulating pipe fittings. 4. The operating voltage should be strictly controlled at 220V±10V. 5. Do not turn on or off the shrink machine during use, turn off the water pump, refrigeration, and power switches in turn after use. 6. After use, release the water in the water bath, and wipe all parts for later use. 7. The compressor is prone to failure when it is not used for a long time, so this equipment should be carried out once a month. Test run again to prevent damage to the compressor.


同類產品推薦


提示

×

*您想獲取產品的資料:

以上可多選,勾選其他,可自行輸入要求

個人信息: